ふ MS-70 MX-50 MF-50 ML-50 1WMPD4000477F
8 MS-70 / MX-50 / MF-50 / ML-50 Caution for Heating (Drying). When the drying temperature is set to 200°C and measurement is started, the t
MS-70 / MX-50 / MF-50 / ML-50 9 3. Outline and Features The moisture analyzer was designed using a super hybrid sensor (S.H.S.) adopted in an
10 MS-70 / MX-50 / MF-50 / ML-50 The analyzer is equipped with a serial interface as standard. It can be connected to a printer or computer.
MS-70 / MX-50 / MF-50 / ML-50 11 4. Packing List and Names of each part Keep the packing box to move the analyzer. Packing list as follows
12 MS-70 / MX-50 / MF-50 / ML-50 Accessory Name Accessory List O : Standard accessory, - : Accessory by your order. M
MS-70 / MX-50 / MF-50 / ML-50 13 4.1. Display and Keys Name State and Meaning At gram display Preset time is displayed at timer mode 1
14 MS-70 / MX-50 / MF-50 / ML-50 #1 ML-50 can select "standard drying " and "quick drying" only. #2 ML-50 does not display &q
MS-70 / MX-50 / MF-50 / ML-50 15 5. Preparations 5.1. Installing the Analyzer 1. Select the place to install the analyzer. Refer to "2
16 MS-70 / MX-50 / MF-50 / ML-50 5.2. Setting the Clock and Calendar Adjust the built-in clock and calendar before use. 5.2.1. Operation
MS-70 / MX-50 / MF-50 / ML-50 17 No good Good 5.3. Proper Operation for Precision Measurement 5.3.1. Operation of the sample Use a proper samp
This manual and Marks All safety messages are identified by the following, “WARNING” or “CAUTION”, of ANSI Z535.4 (American National Standard Ins
18 MS-70 / MX-50 / MF-50 / ML-50 5.3.3. The Glass Fiber Sheets Use the glass fiber sheet to measure the moisture content that is included in t
MS-70 / MX-50 / MF-50 / ML-50 19 6. Measurement Procedure 6.1. Standard Mode Operation The standard mode can obtain the moisture content with
20 MS-70 / MX-50 / MF-50 / ML-50 1. Turn on the analyzer. The gram unit (of weighing mode) is displayed. Enter the Standard Mode 2. Press t
MS-70 / MX-50 / MF-50 / ML-50 21 11. Open the heater cover. Put in a sample using the level indicator. Caution The sample needs to be mor
22 MS-70 / MX-50 / MF-50 / ML-50 6.2. Quick Mode Operation The quick mode can obtain the moisture content with settings of ACCURACY and dryin
MS-70 / MX-50 / MF-50 / ML-50 23 1. Turn on the analyzer. The gram unit (of weighing mode) is displayed. Enter the Quick Mode 2. Press the S
24 MS-70 / MX-50 / MF-50 / ML-50 11. Open the heater cover. Put in a sample using the level indicator. Caution The sample needs to be m
MS-70 / MX-50 / MF-50 / ML-50 25 6.3. Program Number The measurement conditions of all program numbers are set to the standard mode at the fact
26 MS-70 / MX-50 / MF-50 / ML-50 7. Measurement Programs 7.1. List of Measurement Programs There are five analysis modes. Measurement Prog
MS-70 / MX-50 / MF-50 / ML-50 27 7.1.1. ACCURACY of the Standard Mode and Quick Mode ACCURACY of measurement can be set either HI , MID. or
MS-70 / MX-50 / MF-50 / ML-50 1 Contents 1. Safety and Compliance ...3 2. Precaut
28 MS-70 / MX-50 / MF-50 / ML-50 7.1.2. Analyzing mode of the Automatic Mode When the change of moisture content per one minute is less than t
MS-70 / MX-50 / MF-50 / ML-50 29 How to Select quick mode (Extracts) Press the SELECT key during the gram display. Display quc using the ↓ o
30 MS-70 / MX-50 / MF-50 / ML-50 7.2. Procedures to Store a Measurement Program 7.2.1. Standard Drying This explanation uses the following p
MS-70 / MX-50 / MF-50 / ML-50 31 Select an Analysis Mode 5. Press the SELECT key to select a symbol, it will blink. Select an analysis mode using
32 MS-70 / MX-50 / MF-50 / ML-50 Set the Minimum Scale Value of the Gram Display 13. Select gram display with the SELECT key. Select 0.001 [
MS-70 / MX-50 / MF-50 / ML-50 33 7.2.2. Ramp Drying Standard Mode, Automatic Mode or Manual Mode Refer to page 35 for Timer Mode This explanatio
34 MS-70 / MX-50 / MF-50 / ML-50 Select an Analysis Mode 5. Press the SELECT key to select a symbol, it will blink. Select an analysis mode usin
MS-70 / MX-50 / MF-50 / ML-50 35 Store Parameters and Finish the Operation 14. Press the ENTER key to store the new parameters of the measureme
36 MS-70 / MX-50 / MF-50 / ML-50 Select an Analysis Mode 5. Press the SELECT key to select a symbol, it will blink. Select U-t of the timer
MS-70 / MX-50 / MF-50 / ML-50 37 7.2.3. Step Drying Standard Mode, Automatic Mode or Manual Mode Refer to page 39 for Timer Mode This explanatio
2 MS-70 / MX-50 / MF-50 / ML-50 8. Check Function...42 8.1. Self-Check
38 MS-70 / MX-50 / MF-50 / ML-50 Select an Analysis Mode 5. Press the SELECT key to blink a symbol, it will blink. Select an analysis mode using
MS-70 / MX-50 / MF-50 / ML-50 39 Set the Minimum Scale Value of the % Display 13. Select the % display with the SELECT key. Select 0.01 [%] wi
40 MS-70 / MX-50 / MF-50 / ML-50 1. Display gram unit (of the weighing mode). Select a Program Number to Edit the Measurement Program 2. Pres
MS-70 / MX-50 / MF-50 / ML-50 41 Set the Measurement Unit 11. Select measurement unit with the SELECT key. Select the moisture content (based
42 MS-70 / MX-50 / MF-50 / ML-50 8. Check Function 8.1. Self-Check Function Use the self-check function to check whether there is an error
MS-70 / MX-50 / MF-50 / ML-50 43 8.2. Test Sample (Sodium Tartrate Dihydrate) ---Test sample (Sodium tartrate dihydrate, Na2C4H4O6•2H2O )--- S
44 MS-70 / MX-50 / MF-50 / ML-50 9. Connecting to a Printer The analyzer can be connected to a compact dot-matrix printer (AD-8121B) using t
MS-70 / MX-50 / MF-50 / ML-50 45 9.1. Print samples 9.1.1. Example To Print The Whole Data At One Time This example is printed items of "a
46 MS-70 / MX-50 / MF-50 / ML-50 9.1.2. Example To Print Selected Items This print example is printed multiple measurement data and a suit of
MS-70 / MX-50 / MF-50 / ML-50 47 Print Example üüüüüüüüüüüAü&üD Factory MODELüüüüüüMX-50 Product S/NüüüüüP1234567 Serial number IDüüüüüüüLAB-123
MS-70 / MX-50 / MF-50 / ML-50 3 1. Safety and Compliance WARNING Do not use a sample that could make a dangerous chemical reaction and cause a
48 MS-70 / MX-50 / MF-50 / ML-50 9.1.3. Explanation for Print Item "Analyzer information" and "Measurement program" Analysis M
MS-70 / MX-50 / MF-50 / ML-50 49 "Measurement Unit" Unit A Part of Print Formula Display Moisture content is based on wet sample mass 1
50 MS-70 / MX-50 / MF-50 / ML-50 10. Connecting to a Computer The analyzer can be connected to personal computer using the RS-232C interface
MS-70 / MX-50 / MF-50 / ML-50 51 10.1. RS-232C Serial Interface RS-232C Serial Interface Transmission system EIA RS-232C Transmission form
52 MS-70 / MX-50 / MF-50 / ML-50 10.2. Output Format In Case of Format omitted Temperature Data (Function Table 5-d 0 ) The format consists o
MS-70 / MX-50 / MF-50 / ML-50 53 In Case of Format included Temperature Data (Function Table 5-d 1 ) The first 3 figures are the temperature data
54 MS-70 / MX-50 / MF-50 / ML-50 11. Data Memory Function The data memory function automatically stores each result when finishing a measurem
MS-70 / MX-50 / MF-50 / ML-50 55 11.1.2. Output All Data at One Time 1. Display the gram unit (of the weighing mode). 2. Press and hold the
56 MS-70 / MX-50 / MF-50 / ML-50 12. Calibration The moisture content is calculated with a ratio of wet weight and dried weight. Therefore, t
MS-70 / MX-50 / MF-50 / ML-50 57 12.2. Calibration of the Weighting Sensor A standard mass of 20g or 50g can be used for the calibration A
4 MS-70 / MX-50 / MF-50 / ML-50 Compliance with FCC Rules Please note that this device generates, uses and can radiate radio frequency energy.
58 MS-70 / MX-50 / MF-50 / ML-50 Calibration Report Example for the Weighing Sensor Adapted to GLP, GMP and ISO Preset the following parameters to
MS-70 / MX-50 / MF-50 / ML-50 59 12.3. Calibration of Drying Temperature (for MS-70 and MX-50) The temperature calibrator (accessory AX-MX-43)
60 MS-70 / MX-50 / MF-50 / ML-50 10. After fifteen minutes, the buzzer sounds and the display blinks "1"6"0°C . Adjust the bl
MS-70 / MX-50 / MF-50 / ML-50 61 13. Function Table The function table can store the following parameters to control the analyzer. Details of
62 MS-70 / MX-50 / MF-50 / ML-50 13.1.1. Operation 1. Display the gram unit (of the weighing mode). 2. Press and hold the SELECT key to
MS-70 / MX-50 / MF-50 / ML-50 63 14. Maintenance Turn off the power switch and remove power cord during maintenance. Cool down all parts o
64 MS-70 / MX-50 / MF-50 / ML-50 14.2. Replacement of the Halogen Lamp Replace the halogen lamp, when the drying time is excessive or the lam
MS-70 / MX-50 / MF-50 / ML-50 65 14.3. Factory Settings This function can set the following parameters to factory settings. All measurement
66 MS-70 / MX-50 / MF-50 / ML-50 14.4. Troubleshooting 1.In the Case that Proper Results are not be Obtained. Use the self-check function. Re
MS-70 / MX-50 / MF-50 / ML-50 67 14.5. Error Message CH no Internal Error An internal error indicated by the result of the self-check functio
MS-70 / MX-50 / MF-50 / ML-50 5
68 MS-70 / MX-50 / MF-50 / ML-50 15. Specifications MS-70 MX-50 MF-50 ML-50 Measurement method 400 W halogen lamp, thermogravimetric analysis D
MS-70 / MX-50 / MF-50 / ML-50 69 MS-70MX-50MF-50ML-50Power source, Maximum current (r.m.s), Maximum consumptionAC100V to 120V 3A or AC200V to 240V
70 MS-70 / MX-50 / MF-50 / ML-50 15.2. Accessories and Peripheral Equipment Accessories Name Order number Disposable aluminum foil pan (φ83 mm, 1
MS-70 / MX-50 / MF-50 / ML-50 71 16. Index %/min ----------------------------19, 22, 27, 28 , ENTER key ---------------------------14 , PROGRAM key
72 MS-70 / MX-50 / MF-50 / ML-50 I/II------------------------------------------------- 6 id ------------------------------------------------61 ID n
6 MS-70 / MX-50 / MF-50 / ML-50 2. Precautions 2.1. Installing the Analyzer ---Caution for Measurement Safety --- Do not install the analyz
MS-70 / MX-50 / MF-50 / ML-50 7 2.2. During Use ---Caution for Measurement Safety --- Operate the analyzer using the following procedure. P